Content continues after advertisement

From 2013 to 2018, Veterinary Team Brief delivered tools and solutions for the veterinary practice. You are viewing content from the Veterinary Team Brief archive. Find more Veterinary Team Brief content here.

Anesthesia Machines: Finding Leaks

Debbie Coleman, RVT, VTS (Anesthesia), Woodward, Iowa

May 2016|Peer Reviewed|Web-Exclusive

Sign in to Print/View PDF

Anesthesia Machines: Finding Leaks

Benjamin Franklin famously said, “An ounce of prevention is worth a pound of cure”—to address fire safety.1 When a canine or feline patient is waking up during surgery, many reasons (eg, empty vaporizer, hole in the endotracheal tube cuff, hole in the rebreathing bag, cracked sodalime canister) will come to mind and the veterinary technician may feel like he or she is putting out a fire.

Pressure testing the anesthesia machine prior to surgery is “an ounce of prevention.”

The pressure test can help prevent contaminating the team and the environment with anesthetic gases from machine leaks and prevent costly loss of oxygen because of worn gaskets on the E tank yoke. Testing also eliminates the possibility of the patient waking up during the surgery due to a problem with the anesthesia machine.

Related Article: Anesthesia Tips & Tricks for Veterinary Technicians

Conducting a Pressure Test

Perform a pressure test for a rebreathing system as follows:

  • Connect the appropriate circuit and rebreathing bag to the anesthesia machine.
  • Close the pop-off valve.
  • Occlude the y-piece end.
  • Pressurize the circuit with the flush valve and/or flowmeter to approximately 30 cmH2O.2 
  • Turn the flow completely off when the pressure reaches 30 cmH2O and see whether the manometer rises or falls.
  • If the pressure falls but holds with the flowmeter set at 300ml/min or less, then the leak is acceptable, according to FDA standards.3
  • If the leak requires a fresh gas flow rate >300ml/min to hold pressure, locate and correct the leak prior to use.4 
  • If the pressure continues to rise when the flow is completely off, the leak can be either the flowmeter or the flush valve. If the flush valve is leaking the inhalant anesthetic will be diluted, making it difficult to keep patients anesthetized.  If the flowmeter is leaking, the flow will continue and the float will not rest on the bottom of the tube even though the knob is turned off completely.  A service technician can determine the source of leak by using a handheld manometer. Always release the pressure by opening the pop-off valve.

If holding the pressure at 30 cmH2O requires more than 300 ml/min, locate and correct the leak, which can usually be found by spraying the machine with a soapy solution and watching for bubbles.  

Related Article: Anesthetic Considerations for Geriatric Dogs

Finding the Leak

The following scenarios highlight machine areas that should be investigated to pinpoint a leak. 

Scenario 1: The flush valve is used to fill the bag during a pressure test, but the manometer will not rise above 5 cmH2O and falls to zero when the valve is released.

Solution: Ensure the pop-off valve is closed. If so, a leak of this magnitude is typically associated with the gaskets on the sodalime canister. (See Figure 1.) When changing the sodalime, clean the canister and gaskets to ensure a proper seal. (See Figures 2 & 3.) Check the canister’s alignment on the machine.4   Check the pop-off valve by disconnecting the scavenger tubing and plugging the pop-off outlet.  The pop-off can leak because of moisture and a buildup of sodalime dust and will not close completely.

Related Article: Calming Clients' Fears About the Risks of Anesthesia

Scenario 2: The pressure test reveals a 500 ml/min leak.

Solution: Rule out cracks in the sodalime canister, malfunctions with the pop-off valve, pin holes in the rebreathing bag, and cracks and holes in the breathing circuit. Next, check the inspiratory and expiratory domes for cracks and proper alignment. (See Figure 4.) Domes are sealed by O-rings or gaskets; exposed O-rings break down more quickly than those sealed within the machine, whereas gaskets may last longer but eventually become hardened and need to be replaced. Inspect threaded domes for cracks or wear. Replace damaged domes.

Other Leak Sources

The vaporizer is another potential source of leaks. Check the external components by lightly spraying the inlet and outlet end caps and the funnel fill cap with soap solution. Inspect leaking end caps for dents, clean them, and replace on the vaporizer as tightly as possible. Inspect the funnel and the O-ring on the cap of a funnel fill. Both areas can be cleaned with a gauze sponge lightly moistened with alcohol. (See Figure 5.) Use the notch on the cap top to ensure the pin is tight. (Figures 6 & 7.)

Related Article: Quiz: Feline Anesthesia Skills for Technicians

Conclusion

Daily maintenance and proper cleaning of the anesthesia machine will prevent leaks and ensure patient and veterinary team safety. 

References

For global readers, a calculator to convert laboratory values, dosages, and other measurements to SI units can be found here.

Material from Clinician's Brief may not be reproduced, distributed, or used in whole or in part without prior permission of Educational Concepts, LLC. For questions or inquiries please contact us.

Podcasts

Clinician's Brief:
The Podcast

Listen as host Beckie Mossor, RVT, talks with the authors of your favorite Clinician’s Brief articles. Dig deeper and explore the conversations behind the content here.
Clinician's Brief provides relevant diagnostic and treatment information for small animal practitioners. It has been ranked the #1 most essential publication by small animal veterinarians for 9 years.*

*2007-2017 PERQ and Essential Media Studies

© 2018 Educational Concepts, L.L.C. dba Brief Media ™ All Rights Reserved. Privacy Policy (Updated 05/08/2018) Terms of Use (Updated 05/08/2018)